Measurement High Pass Low Pass Test
Radial Vibration Direct
4 Hz
4000 Hz
100 Hz
Velocity Direct
10 Hz
1000 Hz
100 Hz
Acceleration Direct
1000 Hz
10000 Hz
3000 Hz
Hydro Displacement Direct
0.3 Hz
200 Hz
25 Hz
Hydro Velocity Direct
0.3 Hz
200 Hz
25 Hz
Dynamic Pressure
0.5
500 Hz
50 Hz
PP Eccentricity
None
15.6 Hz
1.5 Hz
Recip Rod Runout
None
600 Hz
10 Hz
Aero Velocity Band-pass
3 Hz
3000 Hz
60 Hz
Aero Accel Band-pass (acceleration units)
10 Hz
10000 Hz
200 Hz
Aero Accel Band-pass (integrated velocity units)
25 Hz
350 Hz
100 Hz
Table 22-1) Dynamic Measurement Test Frequencies Examples
Set the function generator amplitude according to the desired measurement level. Set the amplitude
according to the equation:
 = 

1000
For example, if you want the measurement to be 3 mils peak to peak with a 200 mV/mil transducer
setting, set your function generator amplitude to:
3  
200


1000
= 0.6   0.212 
When working with integrated units, use a conversion tool to calculate the non-integrated sensor
signal at the required frequency. There are many internet websites that provide conversion tools.
For example, if you want to test a velocity amplitude of 0.5 in/sec pk at 3,600 rpm, the tool provides a
sensor input of 0.345 g rms. When using a 100 mV/g accelerometer, the input would be 34.5 mV
rms.