EasyManua.ls Logo

BRUEL & KJAER VIBROCONTROL 8000 - 9.1.2 Modbus Serial Port (RS-232, RS-422;485)

BRUEL & KJAER VIBROCONTROL 8000
188 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
© Brüel & Kjaer Vibro S1079330.002 / V04 Page 55 of 188
Technical alterations reserved!
Brüel & Kjaer Vibro Instruction
VC-8000 MPS
Content
EN
9.1.2 Modbus Serial Port (RS-232, RS-422/485)
The SAM provides a single RJ45 connector for Modbus serial
communication. The RJ-45 port is used for convenience; this connector
is not a network port.
The serial port can used to connect to RS-232, RS-422 (4-wire), RS-485
(4-wire), and RS-485 (2-wire) Modbus clients. See Table 6 below for the
RJ-45 port pinout.
An RJ-45 breakout board is used to connect the Modbus client cable(s)
to the serial port (RJ-45) on the rack.
Figure 9-4) Serial Modbus port
RJ-45 PIN RS-485/422 RS-232
Figure 9-5) RJ-45 Breakout board
4
RD-
Rx
5
RD+
-
6
COM
COM
7
TD-
Tx
8
TD+
-
Table 9-2) SAM Modbus (RJ-45) Serial Connector
Figure 9-6) RS-232 and RS-422/485 (4-Wire) wiring examples
Protocol Maximum Cable Length
RS-232
30 m (100 ft.)
RS-422
1200 m (4000 ft.)
RS-485
1200 m (4000 ft.)
Table 9-3) Protocol maximum cable lengths

Table of Contents