Measurement range check
Thisoptionmonitorswhethertheinputsignalremainswithinthevoltagerange
(e.g.±10V).
Wire break monitoring
Thisoptionmonitorswhetherthearithmeticalaverage(DCcomponent)ofthe
inputsignalremainswithinthesetvoltagerange.
Anti-aliasing filter (SW)
Theanti-aliasingfiltersoftwareisadigitallow-passwithacut-offfrequency
around40%ofthesetsamplingfrequency.SelectautomatictoworkwithFFTana-
lysis.
Compression
TheI/Omodulecanbesettocompressmeasurementdatafortransmissionofall
activeanalogueinputs.Refertothetime signaltabinsectionCon-