figuration/settingsforI/Omodule.Thisoptionenablesindividualanalogueinputs
tobeexcludedfromsuchcompression.
Filtersettingstab
Filter
Activatesdigitalfilteringofthetimesignal.Optionsarelow-pass,band-passor
high-passfilters.AButterworthcharacteristicispreset.Filterorderisauto-
maticallysetinternally.Cut-offfrequenciescanbesetatthe1Hzto10000Hz
range.
Using a digital high- or band-pass filter limits the num-
ber of analogue filters (depending on board version) to
1 ... 4. An overview is available in the section on digital
filters.
Integrator
Activatesnumericalintegrationofthetimesignal.Thescalingfactorisfixedata
presetvalueof1000.
Examples of application:
Formeasurementsusingaccelerationsensors,avibrationvelocitysignal(unit
m/s2)isconvertedintoavibrationspeedsignal(unitmm/s).Refertothe"Applic-
ationsusingAMDT"manual.
295
Delphin Technology AG
Dezember 2015