U60 Diagnostic Ultrasound System User Manual Inspection and Maintenance
- 186 -
System Surface Disinfection
The validated disinfecting agents for disinfecting the system are:
Ethanol (75%)
Isopropanol (70%)
To disinfect the system surface:
1. Switch off the system and unplug it.
2. Wear protective gloves to prevent infection.
3. Clean the system prior to disinfection.
4. Prepare the disinfectant solution.
5. Wipe the entire exterior surface of the equipment thoroughly with a soft sterile cloth
dampened with the disinfectant solution. Follow the disinfectant manufacturer's
recommended contact time and mode.
6. After disinfection, wipe off the disinfecting agent with a new sterile cloth dampened
with sterile water.
7. Wipe the system with a dry sterile cloth or leave the system to air dry.
8. Inspect the system to ensure that there is no damage.
NOTE:
After cleaning or disinfecting, check if the device, probe, cable and accessories function
well. If any problem is detected, please contact the manufacturer for service before
reusing them.
Inspect the device, probe, cable and accessories, etc. for any damage.
Power on the system, check if the system boots up successfully without errors
and enters the main display.
After power up, check if the AC power indicator and battery status indicator in
lower right of the screen are as described in Section 5.3.
Check the control panel for any damage.
Inspect probes for cracks, openings, or other damage that could allow liquid
entry.
Connect probes to the system, and check if the main display is as described in
Section 5.3.