System Overview
2-53
Main
transformer
Rectifier
bridge
DC bus
capacitor
IGBT Board
AC
Monoblock
F
Q1
Q2
Q3
Q4
kV Control
board
Drive power
transformer
Resonant
circuit
L
D
C
R
Signals from MCB to KV control board are listed as follows.
Safety line for EXPOSURE_COMMAND
KV set value to KV control board
Signals from KV control board to MCB are listed as follows.
KV too low alarm, lower than 35KV
KV too High alarm, higher than 120KV
Inverter over current alarm
KV unbalance in two bridges of inverter.
KV sample value from KV control board
Test Point
The current feedback of inverter output
The control voltage of KV PI
The control voltage of KV PI with inverter current