Technical Reference
9-9
9.3.3. X-ray tube assembly
Maximum continuous heat dissipation
177W
Tube assembly dimension (length
×
Width
×
Height)
327mm ×132 mm×238 mm
Tube assembly weight
17.5 kg
Maximum heat capacity
900,000HU
Maximum cooling rate
12,500HU/min
Cooling method
Nature cooling
Maximum symmetrical radiation field on the X-axis
100cm from the spot:
Fluoro: 235 mm (diagonal)
Radiography: 240mm (diameter)
9.3.4. Collimator
Input power: DC 5V, 1W
Electrical connection mode: through 3mm, 12pins connector
Mechanical connection mode: cast aluminum bracket, screw fitting
Iris collimator:
Fluoroscopy
Nominal diameter: 23/15/11cm (9”/6”/4.5”)
X-Ray beam continuously adjustable to an area of less than 5×5 cm at the image receptor
Radiography
Nominal diameter circle: 23cm (9”)
Shutter collimator
Rotation: 360°
Operation: Controlled on panel