Calibration
4-63
4.7.3. DAP Calibration:
Formal DAP Calibration:
Must do DAP calibration verification before perform DAP calibraiton.
If Just one of Dose/DAP Coefficient is failed, just calculate the failed coefficient
1. DOSE Rate Measurement:
Put the sensor of Dosimeter on the center of Image Intensifier and set Dosimeter in DOSE Rate in
Gy/sec model. Expose as the KV&mA protocol in Area_B, then input each 6 measurement results in
Area_3.
2. Dose Coef Calculation:
Click Calculate Dose Coef” button, “Dose Coefficientswill be calculated and displayed in Area_II.
3. DAP Coef Calculation:
ClickCalculate DAP Coef” button, “DAP Coefficientswill be calculated and displayed in Area_III.
Click OK, DAP calibration is completed and a dialogue withDAP Calibration Donewill be prompted.