Calibration
4-27
4. When the camera centering verification meets the specification, fasten the alignment screws and
camera secure screws.
5. After fasten secure screws, recheck the alignment result. If fail, redo calibration.
4.6.2. BAFT/UBAT Installation:
BAFT Installation
1. Position the C-arm as Monoblock Up and Image Intensifier down.
2. Put the 9-inch BAFT (Beam Alignment Fast Tool) on top of the Image Intensifier.
3. Orient the tool so that the number 1 and 2 on the BAFT tool is away from the mainframe, number
3 and 4 is on the Mainframe side. (See the picture below)
4. Orient the tool so that the tool sides are aligned with the mounting surface of the Image Intensifier.
1/8 Inch HEX
alignment screws
Screws
parallel