Brivo OEC 715/785/865 Mobile C-Arm X-Ray Product Service Manual
3-10
3.3.3. Unpack Workstation
1. Cut off the binding tie and remove the packing of the workstation. Take out the ramp on standby.
And remove the wrapper on the workstation.
2. Remove all the screws on the wood support on both sides.
3. Remove the screws on both sides as below.
4. Lock the wheels to ensure the wheels to land on the wood support. For Brivo OEC 718/785: lock
all wheels of workstation. For Brivo OEC 865: lock the central control wheels.
5. Via fastening the jackscrew on both sides to lift up the workstation enough to remove the wood
support from pallet on both sides.
For Brivo 865 let central control wheels first land on pallet.
6. Remove the metal support bracket for Workstation, by removing the fixed screws.