Brivo OEC 715/785/865 Mobile C-Arm X-Ray Product Service Manual
5-6
5.2.4. Exposure Management
Select “Exposure ConfigTab on the main screen. “Exposure Configscreen will be displayed.
Restore Default:
Restore Default can restore all setting back to new default value, as above picture.
Once click Restore Default button need follow below instructions to recheck and set all of
the items in Exposure Config”.
Exposure Control:
Enable or disable Auto Termination at 9'30" in one shotfunction. If this function is enable, when the
continuous fluoroscopy time reaches 9’30’’, X-rays will be terminated automatically.
Refer to the below “Exposure Management Matrixtable at the end of this section to set
Fluoro Alarm Control.
Select Filmoption to enable or disable film exposure mode.
Dose Group Setting:
Base on the real power input select corresponding input voltage setting.