Replacement
7-45
7.3.11. Exposure Indicator (5331638)
Perform the procedures below to replace exposure indicator:
1. For Brivo OEC 715/785: Release the 2 screws on the strip cover. Remove the strip cover.
For Brivo OEC 865: remove the Monitor back Cover.
Brivo OEC 715/785
Brivo OEC 865
2. Unplug the cable connected to the indicator. And
then remove the four screws on the sheet metal
support of the indicator.
3. Loose the three screws and then you can
remove the exposure indicator and replace it
with a new one. Assemble the sheet metal
support back onto monitors. Then assemble the
strip covers back.
4. Power on the system and take exposure. Verify
that the exposure indicator lights during
exposure.