Brivo OEC 715/785/865 Mobile C-Arm X-Ray Product Service Manual
4-58
4.7. DAP Accuracy Verification and Calibration
4.7.1. DAP Calibration Verification:
Login Administration interface by Service authority, select “Calibration” Tab. DAP calibration buttons
for system with filter and without filter are displayed on right screen.
System with filter shall performance both calibration DAP Calibration with Filter and
Without Filter.
System without filter shall performance DAP Calibration Without Filter and the “DAP
Calibration with Filter” button is in grey.
Provide appropriate X-ray protection on surface of Image Intensifier in calibration and
verification to protect Image Intensifier.