Brivo OEC 715/785/865 Mobile C-Arm X-Ray Product Service Manual
7-42
7.3.10. Parts in Computer and Peripheral
Perform the procedures below to replace the parts of computer.
1. Power off the system and disconnect Workstation’s AC plug
from the AC receptacle.
2. Remove the rear cover, side covers and front cover of the
Workstation.
3. Unplug the cables from the computer. Loosen the screws of
the computer cover. Lift up and remove the cover.
4. Replace the FRU parts (such as the hard disk, display card, Memory Bank, mother board, image
grabber card and Chassis Power) in the computer.
After the display card is replaced, if the software version is 011601 or higher version, it
is required to reinstall the workstation software (OS and Application Software) refer to
section 5.3.1