Troubleshooting
6-5
6.3.3. Message List
ID
Warning
Description
(Displayed in User Interface)
version is
incompatible
with MCB
Mainframe hardware version
exception. The mainframe
stopped. Please call for service.
flashed wrong version
2. MCB Issue
sync signal is
lost
Camera signal can’t detected.
Please restart the system. If
this issue persists, please call
for service.
abnormal.
2. MCB may be abnormal.
3.Maybe Bundle connection
issue or Bundle cable is
abnormally
after checking
external
watchdog chip
during system
exception.The mainframe
stopped. Please restart the
system. If this issue persists,
please call for service.
temperature
sensor's
feedback SPI
line is works
Tube temperature monitoring
exception. Exposure is
disabled. Please restart the
system. If this issue persists,
please call for service.
temperature
sensor's
feedback line is
disconnected
Tube temperature monitoring
exception. Exposure is
disabled. Please restart the
system. If this issue persists,
please call for service.
1. Check the thermal switch
on the top of monoblock.
2. Check connections and
bundle cable.
3. Check signal to MCB.
temperature
has reached
untouch-
temperature
Xray tube housing overheated.
Exposure is disabled. Please do
not touch the housing. Please
wait for the tube to cool down
before making another
1. Check the thermal switch
on the top of monoblock.
2. Check connections and
bundle cable.
3. Check signal to MCB.
Tube is hot. You may continue
with exposure, but discretion is
advised.
bundle cable.
2. Check OIL_TEMP signal
(temperature sensor inside
monoblock) to MCB.