Brivo OEC 715/785/865 Mobile C-Arm X-Ray Product Service Manual
6-6
ID
Warning
Description
(Displayed in User Interface)
overheating
Tube overheated. Exposure is
disabled. Please wait for the
tube to cool down before
making another exposure.
bundle cable.
2. Check OIL_TEMP signal
(temperature sensor inside
monoblock) to MCB.
Collimator iris stuck. Please
restart the system. If this issue
persists, please call for service.
1.Collimator may be
abnormal.
2.Check cable connections
and conductivity.
3.CCB may be abnormal.
4.MCB may be abnormal.
5.F4 (Fuse +15V) at mother
Board (Backplane) may be
open/close
stuck
Collimator leaves open/close
stuck. Please restart the
system. If this issue persists,
please call for service.
1. Collimator may be
abnormal.
2.Check cable connections
and conductivity.
3. CCB may be abnormal.
4. MCB may be abnormal.
5.F4 (Fuse +15V) at mother
Board (Backplane) may be
stuck
Collimator leaves rotation
stuck. Please restart the
system. If this issue persists,
please call for service.
1. Collimator may be
abnormal.
2.Check cable connections
and conductivity.
3. CCB may be abnormal.
4. MCB may be abnormal.
5.F4 (Fuse +15V) at mother
Board (Backplane) may be
channel is
abnormal
Collimator iris monitoring
exception. Iris can’t open or
close. Please restart the
system. If this issue persists,
MCU IRIS width A/D failure,
MCB may be abnormal
feedback SPI
channel is
abnormal
Collimator leaves open/close
monitoring exception. Leaves
can’t open or close. Please
restart the system. If this issue
persists, please call for service.
failure, MCB may be
abnormal