Calibration
4-69
4.9.2. Input Voltage Calibration:
If the voltage is not in the specification, adjust potential meter R52 to ensure the voltage
between TP19 (red) and TP25 (black) on the PDU board is within the (V2*6.25/V1)±0.01V,
4.9.3. Steering Handle and Wheel Coupling Adjustment
Tools and Test Equipment
FSE standard tools
Torque tool
Marker pen
Operation Steps and Contents
1. Remove rear cover by removing 4 screws