Brivo OEC 715/785/865 Mobile C-Arm X-Ray Product Service Manual
7-74
7.3.22. Switches
Hand switch (P/N: 5085113)
1. Remove the rear and front cover of the Mainframe.
Disconnect the cable from the Hand switch.
2. Cut the bundles that secured the Hand switch cable.
3. Remove the Hand switch and replace a new one.
4. Reinstall Mainframe front and rear covers.
5. Power on the system. Press the hand switch to take
exposure. Verify that the image will display on the left
monitor.
Footswitch (P/N: 5085096/ 5085115)
1. Power off system.
2. Insert the footswitch to the connector.
3. Power on system.
4. Press the footswitch to take exposure.
5. Verify that the image will be displayed on the left monitor.
6. For 5085115, press image save Button can save an image displayed on the left monitor.