Brivo OEC 715/785/865 Mobile C-Arm X-Ray Product Service Manual
7-48
5075006/5075981/5075988(Brivo OEC 715/785)
5085003/5075982(Brivo OEC 865)
6. Unwrap and prepare new camera to be assembled immediately in the next step.
Be careful not to touch the camera lens or IMAGE INTENSIFIER output window and keep
both of them clear without any impurity.
7. Assemble the new Camera on the I.I, and tighten 3 camera fastening screws. Pay attention to the
mounting direction of camera on IMAGE INTENSIFIER as original.
8. Reinstalled upper counter balance.
9. Connect the Camera Head Cables and fix them.
10. Refer to 7.3.2 calibration matrix and chapter 4 to do calibration and checks.
11. After calibration and check, shut down system and reinstall covers removed in above steps.
7.3.14. Image Intensifier with power supply
Perform the procedures below to replace the Image Intensifier with power supply (P/N: 5075929 for Brivo
OEC 715/785, 5075930 for Brivo OEC 865, 5075967 for Brivo OEC 715/785/865 ).
During the Image intensifier replacement procedure, at least 2 man powers are required
to avoid C-arm unbalanced condition and injury.
1. Position the C-Arm for easy access and brake the system, then place the wood block (or equivalent
protection) at the bottom of the system as picture below.