Brivo OEC 715/785/865 Mobile C-Arm X-Ray Product Service Manual
9-6
9.3. Specification of Key Components
9.3.1. X-Ray Generator
Nominal electric power
2.0kw (100kV, 20mA, 0.1s)
Maximum electric power
Normal Fluoro
440W (110kV, 4mA)
High Level Fluoro (HLF)
660W (110kV, 6mA; 55kV, 12mA)
Low dose
220W (110kV, 2mA)
Digital Spot
1760W (110kV, 16mA)
Radiography
2.2 kW (110kV, 20mA) @ 200V, 220V, 230V,
240V;
1.1 kW (110kV, 10mA) @ 100V, 110V, 120V.
High frequency generator
40kHz
Voltage
40kV~110kV
Normal Fluoro
When the X-ray tube current reaches its
highest value of 4mA, the obtainable highest
X-ray tube voltage is 110kV.
High Level Fluoro (HLF)
When the X-ray tube current reaches its
highest value of 12mA, the obtainable highest
X-ray tube voltage is 55kV.
Low dose
When the X-ray tube current reaches its
highest value of 2mA, the obtainable highest
X-ray tube voltage is 110kV.
Digital Spot
When the X-ray tube current reaches its
highest value of 16mA, the obtainable highest
X-ray tube voltage is 110kV.
Radiography:
20mA, stationary @ 200V, 220V, 230V, 240V;
the obtainable highest X-ray tube voltage is
110kV;
10mA, stationary @ 100V, 110V, 120V; the
obtainable highest X-ray tube voltage is 110kV.
Current
Normal Fluoro
0.1 mA~4 mA; when the X-ray tube voltage
reaches its highest value of 110kV, the
obtainable highest X-ray tube current is 4mA.
High Level Fluoro (HLF)
0.2 mA~12 mA; when the X-ray tube voltage