Technical Reference
9-7
reaches its highest value of 110kV, the
obtainable highest X-ray tube current is 6mA.
Low dose
0.1 mA~2 mA; when the X-ray tube voltage
reaches its highest value of 110kV, the
obtainable highest X-ray tube current is 2mA.
Digital Spot
0.2mA~16 mA; when the X-ray tube voltage
reaches its highest value of 110kV, the
obtainable highest X-ray tube current is 16mA.
Radiography
1mAs~80mAs @ 200V, 220V, 230V, 240V;
current: 20mA, stationary; the obtainable
highest X-ray tube voltage is 110kV;
1mAs~40mAs @ 100V, 110V, 120V; current:
10mA, stationary; the obtainable highest X-ray
tube voltage is 110kV.
Radiography
Voltage
40~110kVp
Value of mAs
1.0, 1.3, 1.6, 2.0, 2.5, 3.2, 4.0, 5.0, 6.3, 8.0, 10, 13,
16, 20, 25, 32, 40, 50, 63, 80 mAs @ 200V, 220V,
230V, 240V;
1.0, 1.3, 1.6, 2.0, 2.5, 3.2, 4.0, 5.0, 6.3, 8.0, 10, 13,
16, 20, 25, 32, 40 mAs @ 100V, 110V, 120V;
Pulse Fluoro
1/2/4/8pps
The lowest Current Time Product or the
combination of Loading Factors
40kVp, 1mAs

Table of Contents

Other manuals for GE Brivo OEC 715

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Brivo OEC 715 and is the answer not in the manual?

GE Brivo OEC 715 Specifications

General IconGeneral
BrandGE
ModelBrivo OEC 715
CategoryMedical Equipment
LanguageEnglish

Related product manuals