7.3.2. Calibration Matrix
For the parts below, the calibration procedure or some following action should be done after the replacement. For detail
information, please refer to the replacement procedure.
PN Description
Limiting Field Size Check
Monitor
Brightness and
contrast adjust if needed
PWA, COLLIMATOR CONTROL BOARD
MAINFRAME CONTROL BOARD AND APP SOFTWARE
PWA, FILAMENT DRIVE BOARD
ASM, COLLIMATOR_HL WITH COLLIMATOR CONTROL BOARD
KIT
Component, I.I Power Supply
COMPONENT, I.I POWER SUPPLY_TSBA
GigE CAMERA GE1020A With APP SOFTWARE
GigE CAMERA GE1020B With APP SOFTWARE
GigE CAMERA GE1020C With APP SOFTWARE
FRU,9’’ Image Intensifier for 715,785
FRU,9” Image Intensifier for 865
ASM, 9” IMAGE INTENSIFIER _TSBA
ASM, MONOBLOCK COVER WITH LASER AIMER KIT
ASM, LASER AIMER, 9 INCH ENGLISH
5330055/5085503
5085505/5075503
PWA, POWER DISTRIBUTION UNIT