Brivo OEC 715/785/865 Mobile C-Arm X-ray Product Service Manual
8-4
8.2. PM Checklist
Brivo OEC 715/785/865 PM Checklist
Administrative Information
Medical Facility Information
Field Service Engineer Information
Hospital or Clinic Name:
FSE name:
Employee SSO:
Postal Code/Zip: Contact Person:
Installation Date:
Contact Phone: Department:
Workstation Software Version and Build#:
MCB Software Version and Build#:
Footswitch cable and pins
Electrical Plug and power cord Ground Continuity
Interconnect Cable and Connector
Replace Cell Battery (If needed)
Power On Checks Pass Fail
C-Arm Checks Workstation Checks
Mechanical Movement:
Key Switch:
C-Arm Controls:
Emergency Switch:
Pass Fail
Pass Fail
Pass Fail
Pass Fail
Mechanical Movement :
Workstation Operation:
UPS Battery Performance Check:
System Status Backup:
Pass Fail
Pass Fail
Pass Fail
Yes No
X-ray Generator Accuracy Tests
KV Accuracy (± 7%)
Low Dose:
99kVp@2.0mA: ,
Normal Fluoro:
66kVp@4.0mA: , 110kVp@1.0mA:
HLF:
55kVp@4.0mA(Actual 6.0): ,
110kVp@4.0mA(Actual 4.5): ,(Dose Group: Normal 50 and HLF 50)
110kVp@4.0mA(Actual 6.0): , (Dose Group: Normal 50 and HLF 100)/(Dose Group: Normal 88 and HLF
100)
Digital spot:
55kVp@4.0mA: , 80kVp@4.0mA: , 110kVp@4.0mA:
Radiography:
55 kVp@40mAs : , 80 kVp@40mAs: , 110 kVp@20mAs: