Brivo OEC 715/785/865 Mobile C-Arm X-Ray Product Service Manual
3-60
WARNING
The cassette holder uses friction to hold the cassette.
Verify that cassette is held securely within the cassette holder. Unsecured film cassettes
may fall, injuring patients or personnel.
4. Remove the film cassette by pushing the cassette out of the cassette holder.
5. To remove the cassette holder, open the cassette holder handle and press firmly against the side of
the cassette holder.
The cassette holder is designed to snuggle attach to the image intensifier lip when the
handle is engaged. It may be necessary to strike the cassette holder with the palm of
your hand to dislodge the cassette holder from the image intensifier.
3.7.6. Tube Skin Spacer
Paste the multi language label on Skin Spacer.
Install the skin spacer as the following steps:
Skin Spacer
Small Monoblock Cover
Lock of the skin spacer