Brivo OEC 715/785/865 Mobile C-Arm X-ray Product Service Manual
8-14
Workstation Operation Function Check:
Refer to Refer to 3.6.1 for Workstation operation Function check.
System Status Backup:
Refer to 5.2.6 perform below procedure:
1. Perform integrity check.
2. Backup system Database.
3. Export system setup and administration configuration, if needed.
8.6.4. Computer Cell Battery Check
1. Power off the system.
2. Remove the system rear and front covers.
3. Remove computer side cover.
4. Find the computer cell battery of main board.
5. Check the voltage of the cell battery. Replace cell battery if the test voltage is under DC 2.5V.
6. Contact with GE authorized technical support to get BIOS password.
7. Setup BIOS restore default in “Save & Exit” tab.
8.7. X-ray Generator Accuracy Tests
Below tests are to measure the kVp/mA/mAs accuracy of the X-ray generator system.
8.7.1. kVp Accuracy
Refer to section4.2.1 to perform kVp accuracy check and perform KVp calibration if needed.
8.7.2. mA Accuracy
Refer to section4.3.1 to perform mA accuracy check and perform mA calibration if needed.
8.7.3. mAs Accuracy
Refer to section4.4.1 to perform mAs accuracy check and perform mAs calibration if needed.
If the film mode is turned off in Service Mode by unchecking the check box, mAs accuracy
check is not needed in Periodic Maintenance.

Table of Contents

Other manuals for GE Brivo OEC 715

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Brivo OEC 715 and is the answer not in the manual?

GE Brivo OEC 715 Specifications

General IconGeneral
BrandGE
ModelBrivo OEC 715
CategoryMedical Equipment
LanguageEnglish

Related product manuals