Brivo OEC 715/785/865 Mobile C-Arm X-Ray Product Service Manual
3-70
3.7.11. Mono LCD monitor installation
Refer to Chapter 7 to install the Mono monitor.
3.7.12. Software Option Installation
Refer to Chapter 5 to install software option function. It can be implemented in 3.5.5 System
Configuration together with product model configuration.
3.7.13. Room Interface (RIF)
Room Interface (RIF) is an interface interactive with operation room or some peripheral equipment. The
lamps in operation rooms such as Room In Use indicator or X-Ray On indicator can be connected to
system to warn user, injector can be controlled by room interface if necessary, also the exposure can be
restrained of door status in case of excessive radiation to user, all functions in RIF are not mandatory
when operating.
System provides a HDB15 connector as room interface output/Input to connect each indicators, on the
back of workstation. And door switch work mechanism software setup interface in workstation software.