Brivo OEC 715/785/865 Mobile C-Arm X-Ray Product Service Manual
5-4
Above screen is an example and the software version in it may not be the latest version. Take the actual
show version and build number of the software as standard.
About system ID set or upgrade:
Field engineer can set or upgrade system ID of the system. Current system ID is displayed on the screen
automatically. Fill in the new system ID in the textbox and restart system is OK.
The system ID will be set to blank in manufactory. Field engineer shall set system ID during installation
and make sure it’s the same as the SID in service record system (GIB).
5.2.2. Configure Product
Configure Product by typing:
Input Product Name into Area_1 and input
16 digital option key into Area_2. Click <Add> button in
product configuration area.
Activate software option feature by typing:
Input Feature Name into Area_3 and input
16 digital option key into Area_4. Click <Add> button in
software option area.
Configure Product by import from USB disk:
Insert USB disk with license file into USB port of Workstation, in Medium pull down manual select
corresponding subarea and license file, click <import>, click <OK>.
Activate software option feature from USB disk:
1