Brivo OEC 715/785/865 Mobile C-Arm X-Ray Product Service Manual
5-14
Backup the current database file on this screen. The database file contains patient and exam’s
information. Click “Backup” button to save the current database file. A double confirm screen will pop up.
Click “Yes” button to start backup. The process will complete until “Database backup finished.” message
displays. There are two backup files in the system and used by turns.
Click “Restore…” button to enter the Database Restore screen. The backup files’ name contain the time
stamp. You can choose a database according to the time. Click Restore button and click “Yes” on the
double confirm screen to start the restore process. The restore will complete until “Database restore
finished.” message displays.
Data Integrity Check:
The consistency check function is to check the consistency between the database and image files. Click
Data Integrity Check” button on main screen to enter consistency check screen.
Click “Check” button on the screen, and then click “Yes” button in confirmation screen to start the check
process. The result will be displayed on the frame after the process complete.
Upgrade Software:
This feature is for uninstall the APP software of Brivo OEC 715/785/865 C-Arm system. After uninstalling
the APP software system will go back to Operation System interface. Then insert the APP installation CD,
APP installation Wizard will be invoked automatically.
Detailed software upgrade procedure refer to 5.3.
Reset Admin Password:
This feature is to reset Admin password to original.