Brivo OEC 715/785/865 Mobile C-Arm X-Ray Product Service Manual
9-2
9.1. Overview
The policy of GE Healthcare is one of continual product development and improvement.
For this reason, GE Healthcare reserves the right to change the operating characteristics and
specifications of newer products at any time, without prior notice, and without incurring any obligation
relating to previously manufactured items.
The specifications listed are limited to general performance and physical data. Specifications of optional
equipment provided by other manufacturers are given in the applicable manuals provided with those
options.
9.2. System Specifications
9.2.1. Safety Classification
Class I, Type B, IEC60601-1
Ordinary protection against ingress of water
Degree of Protection:
System: IPX0
Footswitch: IPX8
9.2.2. Environmental Requirements
Operating Conditions:
Temperature: +10°C~+40°C (+50F~+104F)
Humidity: 20%~80% Relative, non-condensing
Atmospheric Pressure: 700hPa~1060hPa
Transportation and Storage:
Temperature: -20°C~+55°C (-4F~+131F)
Humidity: 10~95% Relative, non-condensing
Atmospheric Pressure: 500hPa~1060hPa
9.2.3. Power Requirements
Voltage: AC 100V/110V/120V|200V|220V/230V/240V
Frequency: 50Hz/60Hz
Rated Momentary Line Current (shorter than 5s):
30A (100V/110V/120V)