1.2.15. Lock-Out, Tag-Out
This procedure meets the basic LOTO requirements established by OSHA and follows the policy of GE
Healthcare for energy-isolation during the service/maintenance of this equipment. It is designed to
insure that each employee involved in servicing this equipment has exclusive control over when the
equipment is reenergized (i.e. each employee maintains a personal lock with a unique key).
It is the obligation of each employee to follow this procedure and understand that failing to do so can
cause serious injury or death.
Only Field Service Engineers who have complete LOTO Authorized and initial LOTO Verification training
will be permitted to execute Lockout/Tagout on this system. All other non-authorized individuals will be
considered “affected” individuals.
Specific Source of Energy
Inverter module(Connected with
Capacitors)
Energy Control Procedures
1. Notify all affected individuals of a shutdown.
2. Verify safety of work area.
3. Identify all electrical sources.
4. Position C-arm for service.
5. Shut down system via the power breaker.
6. Remove A/C plug from wall power.
7. Remove interconnect cable from C-arm.
8. Apply the red lock and tag to the AC plug.
9. Remove Workstation back cover.
10. In Workstation, verify zero volts AC at breaker between the brown and blue wires.