Introduction and Safety
 
                       33   
It is applicable for Ed4.0 only 
Portable RF communications equipment (including peripherals such as antenna cables 
and external antennas) should be used no closer than 30 cm (12 inches) to any part of 
this equipment, including cables specified by the manufacturer. Otherwise, degradation 
of the performance of this equipment could result. 
Minimum separation distances for higher IMMUNITY TEST LEVELS shall be calculated using the following 
equation: 
        
 
Where P is the maximum power in W, d is the minimum separation distance in m, and E is the IMMUNITY 
TEST LEVEL in V/m. 
 
 
Use Recommendations 
This product complies with IEC 60601-1-2 EMC standard for medical devices and with radio frequency 
emission requirements per CISPR11 Group1 Class A standard limits. The system is intended for use in 
hospitals. 
Adhering to the distance separation recommendation table, between 150 kHz and 2.5 GHz, will reduce 
disturbances recorded at the image level but may not eliminate all disturbances. 
 However, when installed and operated as specified herein, the system will maintain its essential   
performance. 
 For example, a 1W mobile phone (800 MHz to 2.5 GHz carrier frequency) shall be put 2.3 meters apart 
from the system (in order to avoid image interference risks).