Periodic Maintenance
8-9
8.5.2. UPS performance check:
1. Power on system, waiting for system boot up completely.
2. Double click UPS Display/Menu button, UPS battery capacity
will be displayed on screen. If the displayed capacity is less
than 3 min, the UPS need be replaced.
3. Login software service tool, in log viewer Tab, in Exception
Code input area input 0408018 then click search button. If
code 0408018 is recorded in log, the UPS also need be
replaced.
4. Reassemble computer cover.
5. Reassemble all workstation covers which were removed in
step1.
Replace the UPS battery regularly after being installed for 3/6/9 years.