Brivo OEC 715/785/865 Mobile C-Arm X-Ray Product Service Manual
4-62
Mag
D
iameter (mm)
exposure
mode
kVp
mA
R
eading dose
(mGy)
DAP measured
(Gy.cm2)
DAP on monitor
before exposure
(Gy.cm2)
DAP on monitor
after exposure
(Gy.cm2)
DAP estimated
(Gy.cm2)
Error (%)
Mag2
Normal
fluoro
45 4
Mag2
Normal
fluoro
105 1
9. If system cant pass the verification, perform the Collimator Calibration and DAP calibration, then
perform this verification again, until system passes the verification.