Brivo OEC 715/785/865/865 Mobile C-Arm X-Ray Product Service Manual
2-62
connected means FOOT BUTTON 1 has
been pressed
FOOT_BUTTON2
Button 2
Passive
Open/Close
J1-A19
FOOT_BUTTON2 & FOOT_COM is
connected means FOOT BUTTON 2 has
been pressed
FOOT_COM
Common
termination
Passive
-
J1-B32
-
Emergency switch, located on the control panel of C-arm mainframe, stops all motorized motion immediately,
interrupts any X-rays in progress, and prevents further X-rays until you reset the system. At any time, if the
emergency switch is pushed down, the X-ray exposure shall be stopped immediately. System shall trigger the
related error message.
Name
Description
Attribution
Scale
Comment
EMERGENCY_SWT
One termination of
Emergency
switch1
Passive
Open/Clos
e
EMERGENCY_SWT & REF_EMERGENCY_SWT is
connected means EMERGENCY_SWT has pressed
REF_
EMERGENCY_SWT
The other
termination of
Emergency
switch1
Passive
-
-
EMERGENCY_SWT2
One termination of
Emergency
switch2
Passive
Open/Clos
e
EMERGENCY_SWT2 & REF_EMERGENCY_SWT2 is
connected means EMERGENCY_SWT2 has pressed
REF_
EMERGENCY_SWT2
The other
termination of
Emergency
switch2
Passive
-
-
MF_AC28V_L
-
-
-
-
MF_AC28V_N
-
-
-
-