Brivo OEC 715/785/865/865 Mobile C-Arm X-Ray Product Service Manual
2-70
GigE Camera GE1020A uses an interface flange for mounting itself into image intensifier. This interface
flange is independent part from lens. This interface flange provide three holes and these holes
accommodate errors of at least 2 mm for positioning the image center.
GigE Camera GE1020B use Sysnix HKSC-1 lens. This lens has 16mm focal lens, manual iris F1.6-F11 and
manual focus.
GigE Camera GE1020B uses an interface flange for mounting itself into image intensifier. This interface
flange is integrated into lens. This interface flange provides three holes for mounting camera with image
intensifier, at the same time it provides 6 screws around flange to position the image center for
accommodating errors of at least 2 mm.