Brivo OEC 715/785/865 Mobile C-Arm X-Ray Product Service Manual
3-28
Worklist Check:
Click Press key to enter
Exam Management screen.
Click Schedule…” button to
enter Schedule Exam screen.
In Schedule Exam screen, check
the radio button of Worklist
Server and select target Server,
click Quick Query” button then
patient list in Worklist server will
be displayed if all configurations
and parameters are correct.
6. Storage & Commitment Servers
Setup and Check
Storage & Commitment Servers
setup:
Click “Storage & Commitment
Serversin setup screen.
Click “Add…” button into add
screen. Fill in the information of
the DICOM storage and
commitment server.
Click “Edit…” button into edit
screen. Edit the information of
the existing DICOM storage and
commitment server.
In add and edit screen, click
“Verify” button. The message
“The DICOM server has verified
successfully.” will display if the
server has connected. Click “OK
button and then select the “Exit”
button to close the DICOM
Storage server list screen.
Use Deletebutton to delete
existing DICOM Storage and
commitment server.