Brivo OEC 715/785/865 Mobile C-Arm X-Ray Product Service Manual
3-54
5. Put the video distributor on the backboard and fixed video distributor in workstation cabinet.
Connect video cable firstly, then fasten the 4 secure screws to fix video distributor.
6. Connect the wires to the terminal blocks at the bottom of the workstation (connect the L line to
connector 23 or 24, N line to 25, or 26 and the ground line to earth). Put the power supply cable
along the wiring harness to get a good layout.
7. Plug the video cable of video distributor to the computer’s video connector DVI, which was
connected with the left monitor’s cable before.
5085777)