Brivo OEC 715/785/865 Mobile C-Arm X-Ray Product Service Manual
3-74
Auto Technique Tracking Image Resolution
Copper
Filters
(mm)
Specification (kVp)
Measured
(kVp)
Adjustment
done?
Mag Mode
(Field Size)
Measured
(LP/mm)
Adjustment
done?
(II PN:
5075001)
865 (I.I PN:
5085917)
(I.I PN:
5075008)
718/785 865
Normal Fluoro/Low Dose Fluoro
Dose group: Normal 50 and HLF 50
Dose group: Normal 50 and HLF 100
Dose group: Normal 88 and HLF 100
Camera centering Collimator Centering
Dose error <10% DAP error <10% in calibration
Verification
Brivo OEC 715/785 <6 scales
Dose error <30% DAP error <30% in application
Verification
System Status Backup (If needed)
Function Check for Option
Option Description Verified Not Installed Option Description Verified
Pediatric Application:
Removable Filter and Sensor
Removable Grid
Pedal foot switch:
Dual pedal footswitch with save
button
Yes No
2579 Form Submission ID # # (where applicable)