Calibration
4-11
5. mA accuracy in Radiography Mode:
Measure the voltage of TP7 (MA-SMP) on MCB board with oscilloscope:
mA measured=Sample Voltage (TP7) (unitV)*1000/200ohm
mA real=mA accuracy measure
mA accuracy=[(mA set mA actual )/mA set]*100%
Set the exposure mode and parameter to check to the voltage (TP7) values. The voltage of TP7 should be
in the range of the form below. Otherwise, mA calibration procedures should be performed until all the
voltage values meet the requirement in the list.
For power input: 100V/110V/120V (Setting in Exposure Management)
Exposure mode
KV and mAs value setting
mA value
Voltage (TP7)
minimum (V)
Voltage (TP7)
maximum (V)
Radiography
40KVp@40mAs
10mA
-2.20
-1.80
110KVp@40mAs
10mA
-2.20
-1.80
For power input: 200V/220V/230V/240V (Setting in Exposure Management)
Exposure mode
KV and mAs value setting
mA value
Voltage (TP7)
minimum (V)
Voltage (TP7)
maximum (V)
Radiography
40KVp@80mAs
20mA
-4.40
-3.60
110KVp@40mAs
20mA
-4.40
-3.60