Brivo OEC 715/785/865 Mobile C-Arm X-Ray Product Service Manual
4-30
5. Set to normal mode, fully open the collimator and make a shot.
6. Align the UBAT tool center with image display center.
Ensure UBAT tool is aligned perfectly before perform calibration below.