Calibration
4-33
Cleanup
1. After completing collimator iris centering calibration, make sure to tighten all Collimator Iris
secure screws.
2. Take exposure in Normal Fluoro and ABS mode Recheck the calibration result, if fail, redo the
calibration.
3. Assemble the collimator cover.
4. Remove the BAFT tool from the Image Intensifier
2 Alignment screws