Brivo OEC 715/785/865 Mobile C-Arm X-Ray Product Service Manual
5-8
DAP Unit Config:
Field engineer can configure software DAP unit on this interface.
HLF level exposure setting:
Disable after High Level Exposure: Enable or disable “The system automatically Disable After High Level
Exposure” function. Some countries or regions, such as certain provincial locations in Australia, may
require enabling it.
Disable after No Exposure for 5 Minutes: Enable or disable “Disable after No Exposure for 5 Minutes”
function. Some countries or regions, such as certain provincial locations in Australia, may require
enabling it.
HLF Accessible Only in Auto ABS: Enable or disable “HLF Accessible Only in Auto ABS” function. Some
countries or regions, such as certain provincial locations in Australia, may require enabling it.
Automatic Termination at 20” in One Shot: Enable or disable “Automatic Termination at 20” in One Shot”
function. Some countries or regions, such as Australia or Switzerland, require enabling it.
Refer to the following table “Exposure Management Matrix” to set HLF scene setting.
Exposure Management Matrix
Countries /
Regions
Dose Group Option
HLF
Only in
Auto ABS
Automatic
at 20” in
One Shot
High Level
Exposure
Disable after
No Exposure
for 5
Minutes
Auto
Termination
shot