Troubleshooting
6-15
Exception
ID
Error
Warning
Exception
Description
(In system Log)
Message
(Displayed in User Interface)
Diagnosis Reference
0117005
Warning
KV feedback
channel SPI is
abnormal
Execution exception. Please
restart the system. If this issue
persists, please call for service.
1.MCB may be abnormal
0117006
Warning
MA feedback
channel SPI is
abnormal
Execution exception. Please
restart the system. If this issue
persists, please call for service.
1.MCB may be abnormal
0117007
Warning
Little MA
feedback
channel SPI is
abnormal
Execution exception. Please
restart the system. If this issue
persists, please call for service.
1.MCB may be abnormal
0117008
Warning
KV feedback
channel SPI Tx
time out
Execution exception. Please
restart the system. If this issue
persists, please call for service.
1.MCB may be abnormal
0117009
Warning
KV feedback
channel SPI Rx
time out
Execution exception. Please
restart the system. If this issue
persists, please call for service.
1.MCB may be abnormal
0117010
Warning
Preheat MA
feedback
channel SPI Tx
time out
Execution exception. Please
restart the system. If this issue
persists, please call for service.
1.MCB may be abnormal
0117011
Warning
Preheat MA
feedback
channel SPI Rx
time out
Execution exception. Please
restart the system. If this issue
persists, please call for service.
1.MCB may be abnormal
0117012
Warning
Little MA
feedback
channel SPI Tx
time out
Execution exception. Please
restart the system. If this issue
persists, please call for service.
1.MCB may be abnormal
0117013
Warning
Little MA
feedback
channel SPI Rx
time out
Execution exception. Please
restart the system. If this issue
persists, please call for service.
1.MCB may be abnormal
0117014
Warning
MAS output is
over limit
Execution exception. Please
restart the system. If this issue
persists, please call for service.
1. MCB maybe abnormal
2.mA feedback is abnormal,
recheck mA accuracy.
0117800
Warning
Preheat MA is
not calibrated
System execution exception.
Please try again. If this issue
persists, please call for service.
N/A
0118001
Warning
Xray leakage is
detected when
there is no
exposure
Execution exception. Please
restart the system. If this issue
persists, please call for service.
Maybe Monoblock
grounding issue.