Brivo OEC 715/785/865 Mobile C-Arm X-Ray Product Service Manual
6-44
ID
Warning
Description
(Displayed in User Interface)
configuration
file is empty
Application configuration
loading exception. Please
restart the system. If this issue
persists, please call for service.
2. If not work, re-install
Workstation software.
Re-install workstation
software.
application
failed
Application start failed. Please
restart the system. If this issue
persists, please call for service.
2. If not work, re-install
Workstation software.
application
failed
Application stop failed. Please
restart the system. If this issue
persists, please call for service.
2. If not work, re-install
Workstation software.
failed
Module access failed. Please
restart the system. If this issue
persists, please call for service.
2. If not work, re-install
Workstation software.
module Invoke
corba service
exception
System start failed. Please
restart the system. If this issue
persists, please call for service.
2. If not work, re-install
Workstation software.
3. If not work, re-install OS.
module register
servant
exception
Workstation communication
exception. Please restart the
system. If this issue persists,
please call for service.
2. If not work, re-install
Workstation software.
3. If not work, re-install OS.
property failed
UI locale exception. Please
restart the system. If this issue
persists, please call for service.
2. If not work, re-install
Workstation software.
resource failed
Get text resource failure.
Application display may be
abnormal. Please restart the
system. If this issue persists,
please call for service.
2. If not work, re-install
Workstation software.
saver failed
Screen saver activiation failed.
Please restart the system. If
this issue persists, please call
for service.
2. If not work, re-install
Workstation software.
3. If not work, re-install OS.