Brivo OEC 715/785/865 Mobile C-Arm X-Ray Product Service Manual
7-24
3. Pull up the brake handle and fix the screw below by a
screwdriver, and then rotate the handle clockwise to the
maximum location.
4. Rotate the brake handle counterclockwise to the
maximum, and drop the handle.
5. Repeat step 2 and step 3 until remove the screw below.
6. Replace a new wig-wag brake handle kit with a revers steps.
Horizontal Movement (Cross-arm) Brake (5333085)
1. Make sure the horizontal movement brake is set.
2. Loosen 2 screws from horizontal brake
handle and Remove the handle.
3. Remove the screws of the cross arm cover
and remove the cover.