Replacement
7-31
5. Connect the cables to Monoblock. Boot up the system.
6. Replace Filament Drive Board with the new one. Confirm the switch J7 is set to small focal spot. If
not, configure to small focal spot.
7. Refer to 7.3.2 and chapter 4 to perform calibrations and checks.
8. Shut down the system.
9. Confirm the Jumper is configured to J6 for new Monoblock with Blue Mark.
Thermal Switch cable
Small