EasyManua.ls Logo

GE Brivo OEC 715 - Page 447

GE Brivo OEC 715
551 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Brivo OEC 715/785/865 Mobile C-Arm X-Ray Product Service Manual
7-40
1. Power off system and refer to 7.3.4 remove Workstation back, side, front covers
2. Disconnect the cables related to system interface board
3. Release 4 secure screws under cabinet, and then pull out cabinet from Workstation frame.
4. Release 5 screws which secure System Interface Board and replace a new one.
5. Connect cables and assemble the cabinet.
6. Assemble Workstation front, side covers.
7. Power on system and check the system can boot up functionally.
Mainframe Control Panel Kit (including membrane, display board and support board) (P/N:
5075907)
1. Remove the rear cover of the C-Arm.
2. Disconnect all cables connected to control panel.
3. Loose the screws that secure the control panel and remove the metal shield strips. Then remove
the control panel.
4. Replace a new control panel and connect the cables. Install the C-Arm cover.
5. Power on the system. Pressing the buttons and verify that relative indicator can light.
Workstation Control Panel Kit (5075908)
1. Remove all screws that secure the front upper cover.
Cabinet

Table of Contents

Other manuals for GE Brivo OEC 715

Related product manuals