Brivo OEC 715/785/865 Mobile C-Arm X-Ray Product Service Manual
 
28 
To comply with the regulations applicable to an electromagnetic interface for a Group 1, 
Class A Medical Device, and to minimize interference risks, the following requirements 
shall be applied: 
1.  All cables to peripheral devices must be shielded and properly grounded. The use 
of cables that are not properly shielded and grounded may result in the 
equipment causing radio frequency interference in violation of the European 
Union Medical Device directive and FCC regulations. 
2.  All of the  recommended guidance regarding  electromagnetic environment shall 
be followed. 
 
Guidance and manufacturer’s declaration – Electromagnetic Emissions 
The 
  Mobile C-arm  X-ray  Product  is intended to use in the specified 
electromagnetic environment. The purchaser or operator of the 
Brivo OEC 715/785/865 Mobile 
C-arm X-ray Product should assure that it is used in an electromagnetic environment as described 
below:   
 
Electromagnetic Environment Guidance
 
RF Emissions 
C(Conducted and 
radiated) 
CISPR 11 
Group1 
 
The  Brivo OEC 715/785/865  Mobile C-arm  X-ray 
Product  uses RF energy only for its internal function. 
Therefore, its RF emissions are very low and are not 
likely to cause any interference in nearby electronic 
equipment.
 
lass A 
Brivo  OEC 715/785/865 Mobile C-arm X-ray Product is 
suitable for use in all 
establishments other than 
domestic and those directly connected to  the public 
low-
voltage power supply network that supplies 
buildings used for domestic purposes. 
IEC 61000-3-2 
applicable 
flicker emissions 
IEC 61000-3-3 
applicable