Replacement
 
                   7-65   
 
 
 
Interconnect Cable Workstation 
3.  Remove the four screws securing the casting, and then remove the four screws securing the inner 
socket to take out the head of GigE Ethernet cable. 
      
4.  Replace the new cable and remove the four screws securing the inner socket to assemble the head of 
GigE Ethernet cable. 
5.  Fasten the four screws securing the inner socket and securing screws of casting. 
6.  Connect all the cables and route as original. 
Keep new cable routed as original to ensure the package of mainframe can be open and 
closed smoothly. 
 
7.  Recheck all connectors are correctly connected and cables are well routed. 
8.  Power on the system, and take an exposure. 
9.  If system works well, turn off system and assemble all covers.