Brivo OEC 715/785/865 Mobile C-Arm X-Ray Product Service Manual
 
7-68 
After removing Monoblock, collimator and Image 
Intensifier, C-cam will in unbalance status. The 
back of C-arm will lower than normal. Take care 
anything would be pressured by C-arm. Find wood 
block to support the mainframe is appreciated. 
 
 
9.  Let C-arm in Up/Down position and Down/Up position to make 
the cable go through the C-arm. 
   
10. Install Monoblock support, monoblock 
11.  collimator and Image intensifier. 
12. Connect relative cables, install the new bundle cable sleeves and cable supports, route all the cables 
and install all the other parts which were removed on above steps. 
13. Power on system to ensure system boot up successfully. 
14. Refer to 7.3.2 calibration matrix and chapter 4 to do calibration and checks. 
15. After calibration and check, shut down system and reinstall covers removed in above steps.